Fluke 2635A Power Supply User Manual


 
Performance Testing and Calibration
Introduction
4
4-3
4-1. Introduction
This section of the Service Manual provides performance tests that can be used at any
time to verify that Hydra (2620A, 2625A, or 2635A) operation is within published
specifications. A complete calibration procedure is also included. The performance test
and, if necessary, the calibration procedure can be performed periodically as well as after
service or repair.
4-2. Required Equipment
Equipment required for Performance Testing and Calibration is listed in Table 4-1.
Table 4-1. Recommended Test Equipment
Instrument
Type
Minimum Specifications Recommended
Model
Multifunction
Calibrator
DC Voltage:
Range: 90 mV to 300V dc.
Accuracy: 0.005%
Fluke 5700A
AC Voltage:
Frequency
1 kHz
100 kHz
Voltage
29 mV to 300V
15 mV to 300V
Accuracy
0.05%
1.25%
Frequency:
10 kHz 1V rms .0125%
Decade
Resistance
Source
290 or 190
2.9 k1.9 k
29 k19 k
290 k190 k
2.9 M1.9 M
Accuracy
0.003%
0.003%
0.003%
0.003%
0.0005%
Fluke 5700A
Mercury
Thermometer
0.02 degrees Celsius resolution Princo ASTM-56C
Thermocouple
Probe
Type K Fluke P-20K
Room
Temperature
Oil/Water Bath
Thermos bottle and cap
Multimeter Measures +5V dc. Fluke 77
Signal Generator Sinewave, 0.5 to 1V rms
10 Hz to 5 kHz
Fluke PM5136
Alternate Equipment List
(Minimum specifications are the same as in the Standard Equipment List)
Instrument Type
DMM Calibrator
Function/Signal Generator
Decade Resistance Source
Recommended Model
Fluke 5500A
Fluke PM5193 or Fluke PM5136
Gen Rad 1433H