Fluke 2635A Power Supply User Manual


 
Introduction and Specifications
Specifications
1
1-17
Table 1-3. 2620A/2625A Specifications (cont)
Memory Life 10 years minimum over Operating Temperature range Stores: real-time
clock, set-up configuration, and measurement data
Common Mode Voltage 300V dc or ac rms maximum from any analog input(channel) to earth
provided that channel to channel maximum voltage ratings are observed.
Voltage Ratings Channels 0, 1, and 11 are rated at 300V dc or ac rms maximum from a
channel terminal to earth and from a channel terminal to any other
channel terminal.
Channels 2 to 10 and 12 to 20 are rated at 150V dc or ac rms maximum
from a channel terminal to any other channel terminal within channels 2
to 10 and 12 to 20.
Size 9.3 cm high, 21.6 cm wide, 31.2 cm deep
(3.67 in high, 8.5 in wide, 12.28 in deep)
Weight Net, 2.95 kg (6.5 lbs)
Shipping, 4.0 kg (8.7 lbs)
Power 90 to 264V ac (no switching required), 50 and 60 Hz, 10 VA maximum 9V
dc to 16V dc, 10W maximum
If both sources are applied simultaneously, ac is used if it exceeds
approximately 8.3 times dc.
Automatic switchover occurs between ac and dc without interruption.(At
120V ac the equivalent dc voltage is ~14.5V.)
Standards Complies with IEC 1010, UL 1244 and CSA Bulletin 556B.
Complies with ANSI/ISA-S82.01-1988 and CSA C22.2 No. 231 when
common mode voltages and channel 0, 1, and 11 inputs are restricted to
250V dc or ac rms maximum.
Complies with VDE 0871B when shielded cables are used.
Complies with FCC-15B, at the Class A level when shielded cables are
used.
RS-232-C
Connector: 9 pin male (DB-9P)
Signals: TX, RX, DTR, GND
Modem Control: full duplex
Baud rates: 300, 600, 1200, 2400, 4800, and 9600
Data format: 8 data bits, no parity bit, one stop bit, or
7 data bits, one parity bit (odd or even), one stop bit
Flow control: XON/XOFF
Echo: on/off
2625A Data Storage
Storage 2047 Scans
Each scan includes:
Time stamp
Readings for all defined analog input channels
Status of the four alarm outputs
Status of the eight digital I/O
Totalizer count
Memory Battery-backed static RAM
Memory life: 5 years minimum at 25°C