Remote Interface Operation: You can use the MEASure? or
CONFigure command to select the probe type and thermocouple type.
For example, the following statement configures channel 301 for a
J-type thermocouple measurement.
CONF:TEMP TC,J,(
@301)
You can also use the SENSe command to select the probe type and
thermocouple type. For example, the following statement configures
channel 203 for a J-type thermocouple measurement.
SENS:TEMP:TRAN:TC:TYPE J,(
@203)
The following statements use the SENSe command to set a fixed
reference junction temperature of 40 degrees (always in
°C) on
channel 203.
SENS:TEMP:TRAN:TC:RJUN:TYPE FIXED,(
@203)
SENS:TEMP:TRAN:TC:RJUN 40,(
@203)
The following statement enables the thermocouple check feature on
the specified channels (opens are reported as “+9.90000000E+37”).
SENS:TEMP:TRAN:TC:CHECK ON,(
@203,301)
4
Chapter 4 Features and Functions
Temperature Measurement Configuration
109