Agilent Technologies 34970A Switch User Manual


 
Calibration Errors
The following errors indicate failures that may occur during a calibration.
Refer to the 34970A Service Guide for more information.
701 Cal: security disabled by jumper
The calibration security feature has been disabled with a jumper inside
the instrument. When applicable, this error will occur at power-on to
alert you that the instrument is unsecured.
702 Cal: secured
The instrument is secured against calibration.
703 Cal: invalid secure code
You have entered an invalid calibration security code. You must use the
same security code to unsecure the instrument that was used to secure it,
and vice versa. The security code may contain up to 12 alphanumeric
characters. The first character must be a letter, but the remaining
characters can be letters, numbers, or an underscore ( _ ). You do not have
to use all 12 characters but the first character must always be a letter.
The security code is set to “HP034970” when the instrument is shipped
from the factory.
704 Cal: secure code too long
The security code may contain up to 12 alphanumeric characters.
A security code was received which contained more than 12 characters.
705 Cal: aborted
A calibration in progress is aborted when you turn off the instrument or
send a bus Device Clear.
706 Cal: value out of range
The specified calibration value (CALibration:VALue) is not valid for
the present measurement function and range.
707 Cal: signal measurement out of range
The specified calibration value (CALibration:VALue) does not match
the signal applied to the instrument.
708 Cal: signal frequency out of range
The input signal frequency for an ac calibration does not match the
required input frequency for this calibration.
709 Cal: no cal for this function or range
You cannot perform calibrations for most ac current ranges, the 100 M
resistance range, and period.
6
Chapter 6 Error Messages
Calibration Errors
315