Agilent Technologies 34970A Switch User Manual


 
Alarm Register Commands
See the table on page 284 for the register bit definitions.
STATus:ALARm:CONDition?
Query the condition register in this register group (note that this
condition register uses only bit 4). This is a read-only register and bits
are not cleared when you read the register. A *RST (Factory Reset)
clears the “Queue Empty” bit (bit 4) in the condition register. A query of
this register returns a decimal value which corresponds to the binary-
weighted sum of all bits set in the register.
STATus:ALARm[:EVENt]?
Query the event register in this register group. This is a read-only
register. Once a bit is set, it remains set until cleared by the
STATus:ALARm:EVENt? command or *CLS (clear status) command.
A query of this register returns a decimal value which corresponds to
the binary-weighted sum of all bits set in the register.
STATus:ALARm:ENABle <
enable_value>
STATus:ALARm:ENABle?
Enable bits in the enable register in this register group. The selected
bits are then reported to the Status Byte. A *CLS (clear status) will not
clear the enable register but it does clear all bits in the event register.
A STATus:PRESet clears all bits in the enable register. To enable bits in
the enable register, you must write a decimal value which corresponds to the
binary-weighted sum of the bits you wish to enable in the register.
The :ENABle? query returns a decimal value which corresponds to the
binary-weighted sum of all bits enabled by the STATus:ALARm:ENABle
command.
SYSTem:ALARm?
Read the alarm data from the alarm queue (one alarm event is read
and cleared each time this command is executed). See the diagram on
page 251 for more details on the output format.
5
Chapter 5 Remote Interface Reference
Status System Commands
289