To Connect Wiring to a Module
6 mm
20 AWG
Typical
Channel Number:
Slot Channel
5 Install the module into mainframe.
Wiring Hints...
• For detailed information on each module,
refer to the section starting on page 163.
• To reduce wear on the internal DMM relays,
wire like functions on adjacent channels.
• For information on grounding and shielding,
see page 335.
• The diagrams on the next page show how to
connect wiring to a multiplexer module for
each measurement function.
1 Remove the module cover.
2 Connect wiring to the screw terminals.
4 Replace the module cover.
Cable Tie Wrap
(optional)
3 Route wiring through strain relief.
Chapter 1 Quick Start
To Connect Wiring to a Module
20