• Remote Interface Operation: The following program segment configures
the instrument to scan when an alarm occurs.
TRIG:SOURCE ALARM1 Select the alarm configuration
TRIG:COUNT 2 Sweep the scan list 2 times
CALC:LIM:UPPER 5,(
@103) Set the upper limit
CALC:LIM:UPPER:STATE ON,(
@103) Enable the upper limit
OUTPUT:ALARM1:SOURCE (
@103) Report alarms on Alarm 1
ROUT:MON (
@103) Select monitor channel
ROUT:MON:STATE ON Enable monitoring
INIT Initiate the scan
Note: To stop a scan, send the ABORt command.
4
Chapter 4 Features and Functions
Scanning
85