Agilent Technologies 34970A Switch User Manual


 
Calibration Message
The instrument allows you to store one message in calibration memory
in the mainframe. For example, you can store such information as the
date when the last calibration was performed, the date when the next
calibration is due, the instrument’s serial number, or even the name and
phone number of the person to contact for a new calibration.
You can record a calibration message only from the remote interface
and only when the instrument is unsecured. You can read the
message from either the front-panel or over the remote interface.
You can read the calibration message whether the instrument is
secured or unsecured.
The calibration message may contain up to 40 characters. From the
front panel, you can view 13 characters of the message at a time.
Press to scroll through the text of the message. Press again to
increase the scrolling speed.
Storing a calibration message will overwrite any message previously
stored in memory.
The calibration message is stored in non-volatile memory in the
mainframe, and does not change when power has been off, after a
Factory Reset (*RST command), or after an Instrument Preset
(SYSTem:PRESet command).
Front-Panel Operation:
CAL MESSAGE
Remote Interface Operation: To store the calibration message, send
the following command.
CAL:STRING ’CAL: 06-01-98’
Chapter 4 Features and Functions
Calibration Overview