Agilent Technologies 34970A Switch User Manual


 
34905A/6A Dual 4-Channel RF Multiplexers
These modules consist of two independent 4-to-1 multiplexers.
The channels in each bank are organized in a “tree” structure to provide
high isolation and low
VSWR. Both banks have a common earth ground.
This module does not connect to the internal
DMM. You can connect
your signals directly to the on-board
SMB connectors or to the
SMB-to-BNC cables provided with the module.
NOTES:
The 34905A is used for 50 applications. The 34906A is used for 75applications.
You can close only one channel per bank at a time on these modules; closing one channel
in a bank will open the previously closed channel. One channel in each bank is always
connected to COM.
This module responds only to the CLOSE command (OPEN does not apply). To OPEN a
channel, send the CLOSE command to another channel in the same bank.
12
11
Com1
14
13
22
21
Com2
24
23
Bank Switch
99
Bank Switch
98
Chapter 4 Features and Functions
34905A/6A Dual 4-Channel RF Multiplexers
172