Agilent Technologies 34970A Switch User Manual


 
To Copy a Channel Configuration
After configuring a channel to be included in the scan list, you can
copy that same configuration to other channels in the instrument
(including digital channels on the multifunction module). This feature
makes it easy to configure several channels for the same measurement.
When you copy the configuration from one channel to another, the
following parameters are automatically copied to the new channel:
Measurement configuration
Mx+B scaling configuration
Alarm configuration
Advanced measurement configuration
1 Select the channel to copy the configuration from.
Turn the knob until the desired channel is shown on the right side of
front-panel display. For this example, let’s copy the configuration from
channel 103.
2 Select the copy function.
Use the knob to scroll through the measurement choices until you see
COPY CONFIG. When you press to make your selection, the menu
automatically guides you to the next step.
3 Select the channel to copy the configuration to.
Turn the knob until the desired channel is shown on the right side of
front-panel display. For this example, let’s copy the configuration to
channel 105.
3$67(72
4 Copy the channel configuration to the selected channel.
Note: To copy the same configuration to other channels, repeat this procedure.
1
Chapter 1 Quick Start
To Copy a Channel Configuration
25