MEASure:TEMPerature?
{RTD|FRTD},{85|91|DEF}
[,1[,<
resolution>|MIN|MAX|DEF}]] ,(@<scan_list>)
Configure the specified channels for 2-wire or 4-wire
RTD measurements
and immediately sweep through the scan list one time. Use “85” to
specify
α = 0.00385 or “91” to specify α = 0.00391. Note that this command
also redefines the scan list. The readings are sent directly to the
instrument’s output buffer but the readings are not stored in reading
memory. The default (
DEF) type is “85” (α = 0.00385).
For 4-wire
RTD measurements (FRTD), the instrument automatically
pairs channel n with channel n+10 (34901A) or n+8 (34902A) to provide
the source and sense connections. For example, make the source
connections to the
HI and LO terminals on channel 2 and the sense
connections to the
HI and LO terminals on channel 12. Specify the
paired channel in the lower bank (source) as the scan_list channel.
MEASure:TEMPerature?
{THERmistor},{2252|5000|10000|DEF}
[,1[,<
resolution>|MIN|MAX|DEF}]] ,(@<scan_list>)
Configure the specified channels for thermistor measurements and
immediately sweep through the scan list one time. Note that this command
also redefines the scan list. The readings are sent directly to the
instrument’s output buffer but the readings are not stored in reading
memory. The default (DEF) transducer type is a 5 k
thermistor.
MEASure:VOLTage:DC?
MEASure:VOLTage:AC?
[{<
range>|AUTO|MIN|MAX|DEF}
[,<
resolution>|MIN|MAX|DEF}],] (@<scan_list>)
Configure the specified channels for dc or ac voltage measurements and
immediately sweep through the scan list one time. Note that this command
also redefines the scan list. The readings are sent directly to the
instrument’s output buffer but the readings are not stored in reading
memory. For ac measurements, the resolution is actually fixed at
6
1
2
digits; the resolution parameter only affects the number of digits
shown on the front-panel.