Agilent Technologies 34970A Switch User Manual


 
221 Settings conflict: calculate limit state forced off
If you plan to use scaling on a channel which will also use alarms, be sure
to configure the scaling values first. This error is generated if you attempt
to assign the alarm limits first and the instrument will turn off alarms
and clear the limit values.
222 Settings conflict: module type does not match stored state
Before recalling a stored state, the instrument verifies that the same
module types are installed in each slot. The instrument has detected a
different module type in one or more slots.
223 Settings conflict: trig source changed to IMM
This error is generated if you attempt to set the channel advance source
(ROUTe:CHAN:ADVance:SOURce command) to the same source used for
the scan trigger (TRIGger:SOURce command). The command has been
accepted and executed but the scan trigger source is reset to “IMMediate”.
224 Settings conflict: chan adv source changed to IMM
This error is generated if you attempt to set the scan trigger source
(TRIGger:SOURce command) to the same source used for the channel
advance source (ROUTe:CHAN:ADVance:SOURce command).
The command has been accepted and executed but the channel advance
source is reset to “IMMediate”.
225 Settings conflict: DMM disabled or missing
This command is valid only when the internal
DMM is installed and
enabled. Use the INSTrument:DMM? command to determine the state of
the internal
DMM. For more information, see “Internal DMM Disable”
on page 145.
226 Settings conflict: DMM enabled
When the internal
DMM is enabled, the ROUTe:CHAN:ADVance:SOURce
and ROUTe:CHAN:FWIRe commands are not allowed. Use the
INSTrument:DMM? command to determine the state of the internal
DMM.
For more information, see “Internal DMM Disable” on page 145.
251 Unsupported temperature transducer type
An invalid
RTD or thermistor type has been specified. The following
RTDs are supported: α = 0.00385 (“85”) and α = 0.00391 (“91”).
The following thermistors are supported: 2.2 k
(“2252”), 5 k(“5000”),
and 10 k
(“10000”). Example: CONF:TEMP RTD,1,(@101)
Chapter 6 Error Messages
Instrument Errors
310