Agilent Technologies 34970A Switch User Manual


 
External Scanning In this configuration, the instrument sweeps
through the scan list once each time a low-going
TTL pulse is received
on the rear-panel Ext Trig Input line (pin 6).
You can specify a scan count which sets the number of external
pulses the instrument will accept before terminating the scan.
See “Scan Count” on page 86 for more information.
If the instrument receives an external trigger before it is ready to
accept one, it will buffer one trigger before generating an error.
All readings from the scan are stored in non-volatile memory.
Readings accumulate in memory until the scan is terminated
(until the scan count is reached or until you abort the scan).
Mx+B scaling and alarm limits are applied to measurements during
the scan and all data is stored in non-volatile memory.
Front-Panel Operation:
EXTERNAL SCAN
To initiate the scan, press . The
EXT annunciator turns on as a
reminder that an External Scan is in progress. When a
TTL pulse is
received, the scan starts and readings are stored in memory. To stop
a scan, press and hold .
Remote Interface Operation: The following program segment
configures the instrument for an External Scan.
TRIG:SOURCE EXT Select the external trigger configuration
TRIG:COUNT 2 Sweep the scan list 2 times
INIT Initiate the scan
Note: To stop a scan, send the ABORt command.
Ext Trig Input
Gnd
Ext Trig Connector
0 V
Input
5 V
> 1 µs
4
Chapter 4 Features and Functions
Scanning
83