Agilent Technologies 34970A Switch User Manual


 
ROUTe
:CHANnel:DELay:AUTO {OFF|ON}[,(
@<ch_list>)]
:CHANnel:DELay:AUTO? [(
@<ch_list>)]
Disable or enable an automatic channel delay on the specified channels.
If enabled, the delay is determined by the function, range, integration
time, and ac filter setting (see “Automatic Channel Delays,” on page 89).
Selecting a specific channel delay (using the ROUT:CHAN:DELay command)
disables the automatic channel delay. The :AUTO? query returns the
automatic channel delay setting. Returns “0” (
OFF) or “1” (ON).
ABORt
Abort a scan in progress.
INITiate
Change the state of the triggering system from the “idle” state to the
“wait-for-trigger” state. Scanning will begin when the specified trigger
conditions are satisfied following the receipt of the INITiate command.
The readings are placed in the instrument’s internal reading memory
(up to 50,000 readings can be stored; if memory fills, new readings will
overwrite the first readings stored). Readings are stored in memory until
you are able to retrieve them. Use the FETCh? command to retrieve
reading results.
READ?
Change the state of the triggering system from the “idle” state to the
“wait-for-trigger” state. Scanning will begin when the specified trigger
conditions are satisfied following the receipt of the READ? command.
Readings are then sent immediately to the instrument’s output buffer.
You must then enter the readings into your computer or the instrument
will stop scanning when the output buffer becomes full. Readings are not
stored in the instrument’s internal memory when using READ?.
*TRG
Trigger the instrument from the remote interface.
Chapter 5 Remote Interface Reference
Scanning Overview
230