Agilent Technologies 34970A Switch User Manual


 
ROUTe
:CHANnel:FWIRe {OFF|ON} [,(
@<ch_list>)]
:CHANnel:FWIRe? [(
@<ch_list>)]
This command is valid only when the internal DMM is disabled or
removed from the 34970A.
Configure the list of channels for 4-wire external scanning without the
internal
DMM. When enabled, the instrument automatically pairs
channel n with channel n+10 (34901A) or n+8 (34902A) to provide the
source and sense connections. For example, make the source connections
to the
HI and LO terminals on channel 2 and the sense connections to
the
HI and LO terminals on channel 12. Specify the paired channel in
the lower bank (source) as the scan_list channel.
Note that the ROUT:CLOSe, ROUT:CLOSe:EXCL, and ROUT:OPEN
commands ignore the current ROUT:CHAN:FWIRe setting (if no channels
are in the scan list).
The :FWIR? query returns a “0” (4-wire pairing disabled) or “1” (4-wire
pairing enabled) for each channel in the scan list.
ROUTe:DONE?
Query the status of all relay operations. Returns a “1” when all relay
operations are finished (even during a scan).
SYSTem:CPON {100|200|300|ALL}
Reset the module in the specified slot to its power-on state (CPON means
“card power on”). To reset all three slots, specify ALL.
This is equivalent to pressing from the front panel.