Agilent Technologies 34970A Switch User Manual


 
Remote Interface Operation: The following command retrieves stored
readings from memory (the readings are not erased).
FETCH?
Use the following commands to query the statistics on the readings
stored in memory for a specific channel. These commands do not
remove the data from memory.
CALC:AVER:MIN? (
@305) Minimum reading on channel
CALC:AVER:MIN:TIME? (
@305) Time minimum was logged
CALC:AVER:MAX? (
@305) Maximum reading on channel
CALC:AVER:MAX:TIME? (
@305) Time maximum was logged
CALC:AVER:AVER? (
@305) Average of all readings on channel
CALC:AVER:COUNT? (
@305) Number of readings taken on channel
CALC:AVER:PTPEAK? (
@305) Peak-to-peak (maximum-minimum)
The following command retrieves the last reading taken on channel
301 during a scan.
DATA:LAST? (
@301)
The following command clears the contents of statistics memory for
the selected channel.
CALC:AVER:CLEAR (
@305)
Use the following command to determine the total number of
readings stored in memory (all channels) from the most recent scan.
DATA:POINTS?
The following command reads and clears the specified number of
readings from memory. This allows you to continue a scan without
losing data stored in memory (if memory becomes full, new readings
will overwrite the first readings stored). The specified number of
readings are cleared from memory, starting with the oldest reading.
DATA:REMOVE? 12
Chapter 4 Features and Functions
Scanning
92