Performance Verification
4–2
1780R-Series Service Manual
Table 4–1: Recommended Equipment (Cont.)
Electrical Instrument ExamplePerformance Requirements
NOTE: The 1410 Series generators with standard SPG and TSG modules can be used, but this will not allow all checks and
adjustments to be made.
The 1410 and 1411 Option AB are mainframes with modified TSG7 and TSG11 Color Bar Generators that more accurately control
output amplitudes of the standard 75% amplitude bars.
When preparing this procedure, the modules were configured to provide the following signals at the mainframe J1 to J6 rear-panel
connectors:
J1 Black Burst
J2 Color Bars
J3 Switcher
J4 Linearity
J5 Pulse & Bar
J6 Multiburst
When performing linear waveform distortion, differential phase, and gain checks, the signal from the module was connected
directly to the instrument instead of going through the Switcher.
3. Television Generator Provides accurate test signals for
checking SCH Phase accuracy.
Required signals: Color Bar, Black Burst,
and Field 1 Reference.
TEKTRONIX TSG-170A NTSC Television
Generator or TEKTRONIX TSG-271 PAL
Television Generator.
4. Leveled Sine Wave Generator Flatness 1%, 250 kHz to 50 MHz. The
flatness can be characterized (a chart
made of variations) with the TEKTRONIX
Peak-to-Peak Detector (015-0408-00).
TEKTRONIX SG503 Leveled Sine Wave
Generator installed in a TEKTRONIX
TM500 Series Power Module.
When using the TEKTRONIX SG503 the
following additional accessories are
needed:
Two 50 W coaxial cables, Tektronix Part
No. 012-0482-00.
A 50-to-75 W minimum loss attenuator,
Tektronix Part No. 011-0057-00.
5. Digital (Frequency) Counter Range: 10 kHz to 20 MHz; accuracy,
±0.001%.
TEKTRONIX DC 503A used in a TM500
Series Power Module.
6. Peak-to-Peak Detector Amplifier Tektronix Part No. 015-0408-00 (includes
one 015-0413-00 Peak-to-Peak Detector
Head). Installed in a TEKTRONIX TM500
Series Power Module.
7. Detector Head Two each, Tektronix Part No.
015-0413-00 for use with TEKTRONIX
Peak-to-Peak Detector Amplifier (item 6).
One Detector Head is supplied with the
Peak-to-Peak Detector.
8. Voltmeter (DMM) Range: 0 to 100 VDC; accuracy, ±0.1 %. TEKTRONIX DM501A used in a TEK-
TRONIX TM500 Series Power Module.