Digitizer Local Bus
Commands
The commands used to configure the HP E1429B digitizer for Local bus
data transfers are part of the VINStrument subsystem:
VINStrument
[:CONFigure]
:LBUS
:FEED <
source
>
:MEMory
:INITiate
[:MODE] <
mode
>
:RESet
:SEND
:POINts <
count
>
:AUTO <
mode
>
:TEST
:DATA <
voltage_list
>
Local Bus Transfer
Configurations
The programming sequence used for Local bus transfers depends on the
transfer configuration. The transfer configurations covered in this manual
include:
Single digitizer
post measurement transfers from digitizer memory
real time measurement transfers from the digitizer A/D
Multiple digitizers and serial transfers
post measurement transfers from digitizer memory
real time measurement transfers from the digitizer A/D
Multiple digitizers and interleaved transfers
interleaved real time transfers from the digitizer A/D
The programming sequences for the first two configurations follow. Chapter
2 contains example programs using these sequences.
Note Refer to Appendix D "Local Bus Interleaved Data Transfers" for
programming information on transferring data from multiple digitizers in an
interleaved sequence.