:STORe:AUTO
CALibration[<chan >]:STORe:AUTO <mode > selects whether or not the
calibration constants will be automatically stored when commands like
CALibration:GAIN and CALibration:ZERO complete.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
chan
numeric 1 or 2 none
mode
boolean ON|1|OFF|0 none
Comments Executable when initiated: Yes
Coupled Command: No
Enabling CALibration[<chan >]:STORe:AUTO for either channel enables it for
both channels.
Related commands: CALibration:GAIN, CALibration:SECure:STATe,
CALibration:ZERO, CALibration:STORe
*RST Condition: CALibration:STORe:AUTO ON
Example Turn automatic storage of calibration values off
CAL:STOR:AUTO OFF Disable automatic storage
:VALue
CALibration[<chan >]:VALue <number > specifies the voltage level supplied at
the input. This voltage value is then used in subsequent CALibration:GAIN or
CALibration:ZERO commands. The HP E1429 can not distinguish between a value
which is exactly full scale, and one which is an overload -- both cases would
generate the same measured value. Therefore, the voltage specified for
CALibration:VALue is not allowed to be closer than 10 counts from full scale
(approximately 99.5% of full scale). The following table shows the allowable
CALibration:VALues which are closest to full scale on their respective voltage
ranges.
CALibration[<
chan
>]:STORe:AUTO
208 Command Reference CALibration[<
chan
>] Subsystem