HP (Hewlett-Packard) B Network Router User Manual


 
Coupled command: Yes, VINStrument:CONFigure:LBUS:FEED and
VINStrument:CONFigure:LBUS:SEND:POINts are coupled to this command.
If VINStrument:CONFigure:LBUS:SEND:POINts:AUTO is OFF, the
measurement does not complete normally. After the data has been taken and
transferred, the Local bus is left in a running state. Therefore, it is necessary
to send the ABORt command to each digitizer in the data stream (in a left to
right sequence) after the data is transferred and before proceeding with the
next setup or INITiate. Additionally, any other Local bus devices
(non-digitizers) will need to reset their Local bus chips in a left to right sequence.
See "Local Bus Data Transfers" in Chapter 3 for more information.
Related Commands: VINStrument[:CONFigure]:LBUS:FEED,
VINStrument[:CONFigure]:LBUS:SEND:POINts, ABORt
*RST Condition: ON
Example Setup for multiple transfers of direct A/D data
VINS:LBUS:FEED "CONV:BOTH" Send both channels directly from A/D
converter
VINS:LBUS:SEND:POIN 2; POIN:AUTO OFF
Set up to send end-of-block and
end-of-frame every two readings
[:CONFigure]:TEST:DATA
VINStrument[:CONFigure]:TEST:DATA <voltage_list > configures the
HP E1429B for Local bus testing, and transmits the data given in voltage_list. This
data will be temporarily stored in internal memory before being output at the fastest
possible speed over the Local bus. The data will be stored into alternate channels,
beginning with channel 2. The data is output over the Local bus the same way
(interleaved, with channel 2 data point 1 first). Regardless of their current settings,
VINStrument:CONFigure:LBUS:FEED is set to "MEM:BOTH",
VINStrument:CONFigure:LBUS:MODE is set to GENerate, and
VINStrument:CONFigure:VME:MODE is set to OFF. These changes will remain
in effect after the command has completed.
Parameters The voltage_list must be an IEEE-488.2 definite length block containing values in
16-bit integer format, with the most significant byte being the first byte in a pair (i.e.
Motorola format).
VINStrument[:CONFigure]:TEST:DATA
VINStrument Subsystem Command Reference 291