HP (Hewlett-Packard) B Network Router User Manual


 
:SECure:STATe
CALibration[<chan >]:SECure:STATe <mode >,[<code >] enables or disables
calibration security. Disable the calibration security to calibrate the
HP E1429, change the security code, or change the protected user data.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
chan
numeric 1 or 2 none
mode
boolean OFF|0|ON|1 none
code
character
data
1 to 12 characters none
Comments Executable when initiated: Yes
Coupled Command: No
Either channel may be referenced, enabling/disabling is done to both channels at
once with this command.
The code parameter must be present to disable the security, or it generates error
-109,"Missing parameter". The value supplied must match the currently
programmed security code or it generates error -224,"Illegal parameter value". A
1 second delay will then occur before the HP E1429 executes any subsequent
commands.
To enable security, the code parameter is not required, but is checked if it is
present. If a code is given and is incorrect, error -224, "Illegal parameter value"
will be generated.
Security must be disabled to calibrate the HP E1429, or to use the *PUD
command.
Related commands: CALibration:GAIN, CALibration:ZERO,
CALibration:STORe, CALibration:SECure:CODE, *PUD
*RST Condition: unaffected
Example Disabling calibration security
CAL:SEC:STAT OFF,E1429 Disable security, assuming factory-set
security code
CALibration[<
chan
>]:SECure:STATe
206 Command Reference CALibration[<
chan
>] Subsystem