HP (Hewlett-Packard) B Network Router User Manual


 
Related Commands: TRIGger:STARt:TIMer2, TRIGger:STARt:SOURce
*RST Condition: 5.0E-8 seconds.
PERIOD VALUE TABLE
The following sample periods are available with the digitizer’s internal 20 MHz
reference oscillator.
Multiple Period
(Seconds)
Multiple Period
(Seconds)
Multiple Period
(Seconds)
15.0E-821.0E-742.0E-7
10 5.0E-7 20 1.0E-6 40 2.0E-6
100 5.0E-6 200 1.0E-5 400 2.0E-5
1000 5.0E-5 2000 1.0E-4 4000 2.0E-4
10,000 5.0E-4 20,000 1.0E-3 40,000 2.0E-3
100,000 5.0E-3 200,000 1.0E-2 400,000 2.0E-2
1,000,000 0.050 2,000,000 0.10 4,000,000 0.20
10,000,000 0.50 20,000,000 1.0 40,000,000 2.0
100,000,000 5.0 200,000,000 10.0 400,000,000 20
[:STARt]:TIMer2
TRIGger[:STARt]:TIMer2 <period > specifies the time period between each
sampling event of the post-arm portion of a dual rate measurement. The time period
must be a multiple of the reference oscillator period, with allowable multiples being
1,2,4,10,20,40,........1E8, 2E8, 4E8.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
period
numeric reference period to
reference period * 4E8 |
MINimum | MAXimum
seconds
280 Command Reference TRIGger Subsystem