HP (Hewlett-Packard) B Network Router User Manual


 
DIAGnostic
The DIAGnostic subsystem contains several commands which were developed to
test the instrument at the factory. Several of these commands may prove useful for
trouble shooting or special applications, and so they are written up here.
Subsystem
Syntax
DIAGnostic
:CALibration[<
chan
>]
:CONVerge [query only]
:GAIN
:SENSitivity [query only]
:ZERO
:SENSitivity [query only]
:CHANnel[<
chan
>]
:LABel <
label
>
:FETCh[
<chan >
]
<start_addr >, <count
> [query only]
:MEMory[
<chan >
]
:FILL <
num_segments
>, <
count
>
:ADDresses? [query only]
:PEEK <
address >, <bits
> [query only]
:POKE <
address >, <bits
>
:SGET <
bit
> [query only]
:SPUT <
bit >, <value
>
:TEST? [query only]
:CALibration[<chan >]:CONVerge?
DIAGnostic:CALibration[<chan >]:CONVerge? returns a real array containing
convergance data from the latest CAL:ZERO or CAL:GAIN command. This array
contains pairs of data points, with the first data point in a pair being the average of
the readings taken, and the second data point is the new calibration constant which
will be used on the next convergance iteration. The channel number passed in is
irrelevant, as the data brought back always pertains to the most recently attempted
CAL:GAIN or CAL:ZERO.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel
numeric 1 | 2 none
Comments Executable when initiated: Yes
Coupled Command: No
216 Command Reference DIAGnostic Subsystem