:POINts <
count
>
[SENSe[<chan>]]:SWEep:POINts <count> specifies how many readings will be
taken during each ARM:STARt:COUNt cycle of the trigger system. This command
is the same as (and is coupled to) TRIGger:STARt:COUNt. Changing either
changes the value of the other as well.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
chan
numeric 1|2 none
count
numeric 1 | 7 through 16,777,215 |
MINimum|MAXimum|
9.9E+37|INFinity
none
MINimum selects 1 reading.
Memory, post-arm readings only: MAXimum = 524,288 / ARM:STARt:COUNt
Memory battery enabled: MAXimum = (524,288 / ARM:STARt:COUNt) - 4
A/D converter to VME (VXI data transfer) bus: MAXimum = 16,777,215
A/D converter to Local bus: MAXimum = 16,777,215
Digitizer memory, pre- and post-arm readings: see below
ARM:STARt:COUNt Number of
Memory
Segments
Maximum Readings
(TRIGger:STARt:COUNt)
1 1 524,288
2 2 262,144
3 - 4 4 131,072
5 - 8 8 65,536
9 - 16 16 32,768
17 - 32 32 16,384
33 - 64 64 8,192
65 - 128 128 4096
NOTE: If the non-volatile mode of memory is enabled
(MEMory:BATTery:STATe ON), then all of the maximum reading counts
shown above decrease by four. These four memory locations in each
segment hold the data necessary to recover all readings after a power
failure.
[SENSe[<
chan
>]]:SWEep:POINts <
count
>
260 Command Reference [SENSe[<
chan
>]]:SWEep Subsystem