HP (Hewlett-Packard) B Network Router User Manual


 
Code Message Description
-410 Query INTERRUPTED The digitizer was sent a command before it was
finished responding to a query command.
-420 Query UNTERMINATED The controller (computer) attempts to read a query
response from the digitizer without having first sent a
complete query command.
-430 Query DEADLOCKED The digitizer’s input and output buffers are full and the
digitizer cannot continue.
-440 Query UNTERMINATED after
indefinite response
Occurs when the *IDN? query is not the last query
executed in a command string.
1002 Cal security enabled Calibration security must be disabled to calibrate the
digitizer, to read or write calibration data, to change
the security code, or to change the protected user
data.
1004 Cal write fail Writing calibration or protected user data (*PUD) to
nonvolatile memory failed.
1005 Error during CAL An error occurred during calibration. Refer to the
statement appended to this message for a description
of the error.
1007 Calibration security defeated A jumper was moved to defeat calibration security.
1008 Error during zero cal An error occurred during calibration of the zero offset.
Refer to the statement appended to this message for
a description of the error.
1009 Error during gain cal An error occurred during gain calibration. Refer to the
statement appended to this message for a description
of the error.
1010 Error during linearity cal An error occurred during linearity calibration. Refer to
the statement appended to this message for a
description of the error.
1015 A/D control register not responding The serial interface register was not working properly
at power-on.
1016 Illegal during LBUS or VME
memory transfer
The command can not be executed while a VME bus
or Local bus data transfer is in progress.
1017 Battery too low, data may be lost The battery does not contain sufficient charge to
maintain memory over an extended period. This error
occurs when readings are taken, when the battery is
enabled/disabled, or during the self-test.
1018 Battery backed data corrupt This error is due to a low battery charge, or if the
battery is enabled after readings are in memory.
Table B-3. HP E1429A/B Error Messages (Cont’d)
Appendix B Useful Tables 339