HP (Hewlett-Packard) B Network Router User Manual


 
[SENSe[<
chan
>]]:ROSCillator
The SENSe:ROSCillator subsystem controls the reference oscillator source and
frequency used to generate sample rates for taking measurements.
Subsystem
Syntax
[SENSe[<
chan
>]]
:ROSCillator
:EXTernal
:FREQuency <
frequency
>
:SOURce <
source
>
Since the triggering and timebase circuits of the HP E1429 are shared between both
sensing channels, the settings for [SENSe[<chan>]]:ROSCillator are the same for
both values of chan (1 and 2). Setting SENSe1:ROSCillator values will also set
SENSe2:ROSCillator values and vice versa.
:EXTernal:FREQuency
[SENSe[<chan>]]:ROSCillator:EXTernal:FREQuency <frequency> indicates to
the HP E1429 the frequency of an external reference oscillator signal. The
SENSe:FREQuency subsystem uses this value to generate the sample rate when
SENSe:ROSCillator:SOURce is set to EXTernal, ECLTrg0, or ECLTrg1.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
chan
numeric 1|2 none
frequency
numeric 10 KHz through 20 MHz
MINimum|MAXimum
Hz
MINimum selects 10 KHz; MAXimum selects 20 MHz.
Comments Executable when initiated: No
Coupled command: Yes. This command is coupled to the commands in the
TRIGger subsystem.
Indicating an incorrect frequency for an external reference oscillator will cause
the sample rate and trigger delay to be incorrect.
The amplitude of the external reference signal should be a TTL level:
low = 0.0V to 0.8V, high = 2.5V to 5.0V.
254 Command Reference [SENSe[<
chan