HP (Hewlett-Packard) B Network Router User Manual


 
Comments Executable when initiated: No
Coupled command: No
The HP E1429B Local bus chip must be reset after each data transfer. When
resetting the Local bus chip, the Local bus chips on all devices to the right of the
HP E1429B must also be reset in a left-to-right sequence. Refer to the product
documentation for information on how a particular device’s Local bus chip is
reset.
Related Commands: *RST, ABORt
*RST Condition: none
Example Reset the Local bus chip
VINS:LBUS:RES Reset the chip
[:CONFigure]:LBUS:SEND:POINts
VINStrument[:CONFigure]:LBUS:SEND:POINts<count > specifies how many
readings will be output over the Local bus per block. There are two possible count
settings: 1 point per feed channel or all points per feed channel. Normally, there is
no need to set a count with this command because with
VINStrument:CONFigure:LBUS:SEND:POINts:AUTO ON (the power-on and
*RST setting), the number of points will automatically be matched to the current
setting of VINStrument:CONFigure:LBUS:FEED. Therefore, the count will
normally be TRIGger:COUNt * 2 if VINS:LBUS:FEED is "xxx:BOTH" or
TRIGger:COUNt * 1 if the feed is "xxx:CHAN1" or "xxx:CHAN2".
If VINStrument:CONFigure:LBUS:SEND:POINts:AUTO is set to OFF, then the
number of points sent per block must either be 2 ( for feed "CONV:BOTH") or 1
(for feeds "CONV:CHAN1 and "CONV:CHAN2"). This setting is only allowed
when the feed is one of the "CONV:xxx" (direct from A/D converter) settings. The
combination of VINStrument:CONFigure:LBUS:SEND:POINts:AUTO OFF and
VINStrument:CONFigure:LBUS:SEND:POINts 1 or 2 is needed only when the
goal is to multiplex readings directly from multiple digitizers onto the Local bus.
Parameters MINimum and MAXimum will both set the same value, which is the number of
readings that will be transferred as determined by the current
VINStrument:CONFigure:LBUS:SEND:POINts:AUTO and
VINStrument:CONFigure:LBUS:FEED settings.
Comments Executable when initiated: No
VINStrument[:CONFigure]:LBUS:SEND:POINts
VINStrument Subsystem Command Reference 289