MINimum and MAXimum are not allowed with this command.
Optional parameters that are left blank are filled from left to right. Therefore, it
may be necessary to use the syntax DEFault to note that a particular parameter has
been defaulted. For example, to default the number of readings and specify a
sample rate, the command would appear as:
CAL:ZERO DEF, .05, one
CALibration:ZERO forces the internal reference (20 MHz) oscillator to be used.
Sample rates are attained using that reference.
The default number of readings is 1000, and the default period is 1.0E-4 seconds.
These numbers were chosen such that the product of the two is a period that is an
integral multiple of both 50 Hz and 60 Hz line cycles
(0.1 seconds in this case).
The product of the period and number of readings will be checked to see if it
exceeds 10 seconds, and if so, error -221;"Settings conflict; Calibration time too
long" occurs.
The default <mode> is ONE, which calibrates using the current settings of
SENSe:VOLTage:RANGe and SENSe:FUNCtion. Specifying <mode> ALL will
do a zero calibration on all voltage range settings for both of the ports on the
specified channel.
Related commands: CALibration:GAIN, CALibration:VALue,
CALibration:SECure:STATe
*RST Condition: none
Example Performing a zero calibration
CAL:SEC:STAT OFF,E1429 Disable security, assuming factory-set
security code
CAL1:ZERO DEF,DEF,ALL Calibrate channel 1 using default sample
rate and number of points, calibrate all gain
ranges on both ports.
CALibration[<
chan
>]:ZERO
CALibration[<
chan
>] Subsystem Command Reference 211