HP (Hewlett-Packard) B Network Router User Manual


 
The MEASure command configures the HP E1429 to do the measurement
specified by the parameters given with MEASure. All instrument settings will be
forced to their *RST values, with the following exceptions. In particular, note that
all OUTPut signals are turned off and their FEED’s are changed to reset values,
and that both the local bus (VINStrument:LBUS) and VME bus (VINS:VME) are
turned off. Be aware that these new states will still be in effect after the MEASure
command is complete. See the -- mumble which section -- for a complete list of
the reset settings for the HP E1429. The following states are set up after the reset,
and thus in most cases, will not be the *RST value for that command:
a. SENSe<n>:FUNCtion is set to "VOLTage[:DC] <port>"
b. SENSe<n>:VOLTage:DC:RANGe is set to the value implied by the
expected value given.
c. INPut[<port>]:FILTer is set to ON.
d. SENSe:SWEep:POINts and TRIGger:STARt:COUNt are set to the <size>
parameter.
Related commands: SENSe:FUNCtion, SENSe:VOLTage:DC:RANGe,
SENSe:SWEep:POINts, SENSe:ROSCillator:SOURce), TRIGger:STARt
subsystem, ARM:STARt subsystem.
*RST Condition: none
Example Measuring 1.5 V on the differential input of channel 1
MEAS1:ARRay:VOLT? (20),1.5,(@3) Set up, take, and bring back 20 readings of
1.5 Volts peak on port 3 of channel 1.
MEASure[<
chan
>]:ARRay[:VOLTage][:DC]?
MEASure[<
chan
>] Subsystem Command Reference 239