HP (Hewlett-Packard) B Network Router User Manual


 
Readings may not be obtained from memory until the trigger system has returned to
the idle state. The FETCh? and READ? commands may be invoked before the
trigger system is in the idle state. These commands will wait until the trigger
system is idle before returning readings. The trigger idle state may be determined
externally from the high-to-low transition of the BUSY bit (bit 8) of the
STATus:OPERation:CONDition register.
Comments Executable when initiated: No
Coupled Command: No
Both measurement channels are affected by this command.
If the trigger system is not in the idle state, error -123,"Init ignored" will be
generated, and the trigger system will be unaffected.
The ABORt command may be used to prematurely halt the trigger system and
place the HP E1429A in the idle state.
Related Commands: *OPC, *OPC?, *RST, *WAI, ABORt, ARM subsystem,
STATus:OPC:INITiate, TRIGger subsystem
*RST Condition: The trigger system is in the idle state.
Example Placing the HP E1429A in the wait-for-arm state
INIT Initiate signal measurement
INITiate[:IMMediate]
INITiate Subsystem Command Reference 231