Maximum
Expected
Value (V)
Voltage
Range
(Volts)
Resolution
(Volts)
± .1 0.10235 .00005
± .2 0.2047 .00010
± .5 0.51175 .00025
± 1 1.0235 .0005
± 2 2.047 .0010
± 5 5.1175 .0025
± 10 10.235 .005
± 20 20.47 .010
± 50 51.175 .025
±100 102.35 .05
Comments • Executable when initiated: No
• Coupled Command: No
• If no input port is given, CONFigure1 defaults to (@1) and CONFigure2 defaults
to (@2).
• Resolution varies by range, and is constant for each range. For any given range,
specifying MINimum, MAXimum, or DEFault for resolution yields the same
result.
• It is important to note that the expected value determines the resolution and not
vice versa. The digitizer always uses 12-bit resolution and a coarser resolution
value has no effect. If, for some reason a coarser resolution is desired, specify a
larger expected value. See the previous table for expected values and resolutions.
CONFigure[<
chan
>]:ARRay:[VOLTage][:DC]
214 Command Reference CONFigure[<
chan