HP (Hewlett-Packard) B Network Router User Manual


 
Example Selecting the 102.35 Volt range on channel 2
SENS2:FUNC “VOLT4” Connect differential port 4 to channel 2
SENS2:VOLT:RANG 75 Select 102.35 volt range. Any number
greater than 51.175 forces the next highest
(102.35 V) range.
:RESolution?
[SENSe[<chan >]]:VOLTage[:DC]:RESolution? returns the resolution associated
with the current SENS:VOLT:RANGe setting, on the specified channel.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
chan
numeric 1|2 none
Comments Executable when initiated: Yes
Coupled Command: No
*RST Condition: None
Example Querying the current resolution on channel 2
SENS2:VOLT:RES? Query channel 2 for resolution setting
[SENSe[<
chan
>]]:VOLTage[:DC] :RESolution?
264 Command Reference [SENSe[<
chan
>]]:VOLTage[:DC] Subsystem