HP (Hewlett-Packard) B Network Router User Manual


 
Note These pulses will also be generated when data is read out of memory from the
VINStrument:LBUS or VINStrument:VME subsystems.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
n
numeric 0|1 none
source
string “EXTernal[1]”|
“[SENSe[1|2]]:ROSCillator”|
“TRIGger[:STARt|:SEQuence[1]]”
none
Comments Executable when initiated: No
Coupled Command: No
Note that if the feed is EXTernal1, then OUTPut:EXTernal1:STATe must be ON
in addition to OUTPut:ECLTrg<n>:STATe ON for an output to occur on the
chosen trigger line.
The source of the synchronization pulse is independent for each ECLTRG trigger
line.
Related Commands: OUTPut:ECLTrg<n>:STATe, OUTPut:EXTernal1:FEED,
OUTPut:EXTernal1:STATe
*RST Condition:
OUTPut:ECLTrg0:FEED “TRIGger:STARt|SEQuence1”,
OUTPut:ECLTrg1:FEED “EXTernal1”
Example Setting the ECLTrg0 sync pulse source
OUTP:ECLT0:FEED “TRIG” Output a pulse whenever a reading is taken
:ECLTrg<
n
>[:STATe]
OUTPut:ECLTrg<n>[:STATe] <mode> enables or disables the routing of the
selected synchronization pulse to the specified VXIbus ECLTRG trigger line
(ECLTrg0 or ECLTrg1).
OUTPut:ECLTrg<
n
>[:STATe]
OUTPut Subsystem Command Reference 243