HP (Hewlett-Packard) 4 Series Calculator User Manual


 
Chapter 3 Peak Detection and Labeling
3-62 Applied Biosystems
3
Figure 3-23 User Label Setup Dialog Box
4. Select the Label Type (spectra only):
MassLabels with Apex or Centroid mass.
Mass difference from adjacent peaksApplies
the labels to peaks that have the specified mass
difference relative to the adjacent labeled peak of
lower m/z.
5. Select Match Charge State (spectra only) if you want the
charge state of a peak evaluated before applying the user
label. When this function is enabled, a peak must occur
within the specified Mass Tolerance (described below),
and have the specified charge state before the user label
is applied.
Hint: The Match Charge State function allows you to
screen out peaks that are within the specified Mass
Tolerance, but are not the charged species you are
interested in.
Column
buttons
header