HP (Hewlett-Packard) 4 Series Calculator User Manual


 
Chapter 8 Viewing Voyager PSD Data
8-10 Applied Biosystems
8
8.3 Calibrating a PSD Spectrum
NOTE: Multi-point calibration yields higher mass accuracy
than one-point calibration.
This section includes:
Checking peak detection
Calibrating
Creating PSD .CAL files and applying to other data files
Creating PSD calibration reference (.REF) files
Changing the precursor mass
When to use this
procedure
Use this procedure to:
Generate a PSD calibration (.CAL) file from a known
standard analyzed in PSD mode. You can then export the
PSD .CAL file and include the PSD .CAL file in a PSD
instrument settings (.BIC) file you use to acquire unknown
PSD samples.
Internally calibrate an unknown PSD data file by
specifying known monoamino acid fragment ions as the
peaks to match during calibration.
Overview of
creating a PSD
.CAL file
To create a PSD .CAL file:
Acquire a peptide standard with a known sequence in
PSD mode.
Check peak detection as described in Section 8.3.1,
Checking Peak Detection.
Create a calibration reference (.REF) file as described in
Section 8.3.4, Creating PSD
Calibration Reference (.REF) Files.
Internally calibrate the PSD .DAT file as described in
Section 8.3.2, Calibrating. Specify the calibration
reference (.REF) file created in Section 8.3.4, Creating
PSD Calibration Reference (.REF) Files, when
calibrating.
Export the calibration (.CAL) file from the PSD .DAT file to
use when acquiring unknown samples in PSD mode.