HP (Hewlett-Packard) 4 Series Calculator User Manual


 
Chapter 5 Examining Spectrum Data
5-10 Applied Biosystems
5
6. Select the Peak Weighting Factor. If the calibration
includes more than two points, you can apply the following
weighting factors to fit points to the curve:
NoneAll peaks weighted equally
Inverse WidthNarrower peaks are weighted
more than broader peaks
HeightMore intense peaks are weighted more
than less intense peaks
Manually matching
peaks
7. To manually select the reference mass for a peak,
right-click-drag over the peak of interest.
The Select or Create Reference Peak Information
dialog box (Figure 5-3) is displayed and lists all masses
in the selected reference file. The selected entry is the
nearest match in the calibration reference file to the
selected peak.
Figure 5-3 Select or Create Reference Peak
Information Dialog Box