HP (Hewlett-Packard) 4 Series Calculator User Manual


 
Manual Calibration
Data Explorer
Software Users Guide 5-17
5
5. To save the calibration to the data file, select Mass
Calibration from the Process menu, then:
5.3.3 Creating or Modifying a
Calibration Reference File (.REF)
This section includes:
Definition
.REF files provided
Calibration reference file contents
Creating and saving a calibration reference file
Modifying a calibration reference file
Specifying mass type for highly charged, narrow peaks
If you are
calibrating
Select The following occurs
Mariner
data
(including
MS Method
data)
Apply Calibration All spectra in the data file are calibrated and
displayed with an MC trace label. The
calibration constants are saved with the data
file. Each spectrum in the data file is
calibrated when displayed.
Voyager
data
Apply Calibration The current spectrum is calibrated and
displayed with an MC trace label. The
calibration constants are saved with the
spectrum.
Apply to All
NOTE: This button is
displayed only if you
are calibrating a
Voyager
multispectrum data
file.
All spectra in the data file are calibrated
using the currently displayed calibration, and
are displayed with an MC trace label. The
calibration constants are saved with the data
file. Each spectrum in the data file is
calibrated when displayed.