HP (Hewlett-Packard) 4 Series Calculator User Manual


 
General Troubleshooting
Data Explorer
Software Users Guide 9-5
9
Spectra labeled with
spectrum numbers that do
not correspond to the axis
You are viewing
event-filtered MS Method
data. Spectra in an
event-filtered trace are
numbered contiguously
(1,2,3...) regardless of
their relation to the overall
acquisition. However, the
axis of the trace reflects
the numbering of the
overall experiment.
No action. Normal
occurrence.
Conversion of .SPC to
.DAT failed
A .DAT file with the same
file name you are
specifying for the
converted .SPC file is
open
1. Close the open .DAT
file.
2. Use a different name
for the converted .SPC
file to prevent the new
conversion from
overwriting the existing
file.
Table 9-2 General TroubleshootingMariner Only (Continued)
Symptom Possible Cause Action
Table 9-3 General TroubleshootingVoyager Only
Symptom Possible Cause Action
Failed to get original
calibration message
displayed when you open
a data file
You are opening a
Voyager .MSF data file
collected with version 4.0
or earlier software.
No action. Normal
occurrence.