Tektronix cts710 Surge Protector User Manual


 
Setting SDHSONET Transmit Parameters
CTS 710 SONET Test Set User Manual
3–79
Z2 byte
selected to
edit
Bytes you can
never edit
Cannot edit
because
alarm is set
Figure 3–28: Editing the Z2 Transmit Overhead Byte
There are several ways to edit the byte:
H Select Reset Overhead to set all bytes in the overhead to
their individual default values. However, Reset Overhead
does not remove simulated error conditions that have been
set using controls in the ERRORS & ALARMS page.
A simulated error condition overrides the default value of the
overhead byte that signals that particular error condition.
H Select Default (XXXXXXXX) to set the selected byte to its
default value. The specific default value depends on the
selected byte and on the rate and structure of the signal.
H Select EDIT BYTE to define the byte as a value different
from the preset choices.
4. To edit the selected byte to a value of your choice:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
EDIT BYTE
Setting SDHSONET Transmit Parameters
CTS 710 SONET Test Set User Manual
3–79
Z2 byte
selected to
edit
Bytes you can
never edit
Cannot edit
because
alarm is set
Figure 3–28: Editing the Z2 Transmit Overhead Byte
There are several ways to edit the byte:
H Select Reset Overhead to set all bytes in the overhead to
their individual default values. However, Reset Overhead
does not remove simulated error conditions that have been
set using controls in the ERRORS & ALARMS page.
A simulated error condition overrides the default value of the
overhead byte that signals that particular error condition.
H Select Default (XXXXXXXX) to set the selected byte to its
default value. The specific default value depends on the
selected byte and on the rate and structure of the signal.
H Select EDIT BYTE to define the byte as a value different
from the preset choices.
4. To edit the selected byte to a value of your choice:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
EDIT BYTE