Tektronix cts710 Surge Protector User Manual


 
Appendix E: Incoming Inspection Test
CTS 710 SONET Test Set User Manual
E–5
2. Perform the initial setup of the CTS 710 with the following
sequence:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS RECALL
INSTRUMENT
SETUPS
DEFAULT
FACTORY
SETTINGS
Recall Setup
TRANSMIT TRANSMIT
SETTINGS
Tx/Rx Settings Coupled
POINTERS &
TIMING
Pointer Control Set Value
Pointer Value set
to
Default 522
3. To check the STS-1 rate, press the CLEAR HISTORY button,
wait two seconds, and then verify that no yellow history lights are
on.
4. To check the STS-3 rate, set the transmit rate as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TRANSMIT TRANSMIT
SETTINGS
Transmit Rate STS-3
NOTE. When changing the transmit rate, the LOP, LOS, and LOF
history lights may turn on; this is normal. However, the red event
lights should not stay on.
5. Press the CLEAR HISTORY button, wait two seconds, and then
verify that no yellow history lights are on.
6. Steps 7 through 10 of this procedure apply only if one of the
Electrical/Optical Plug-in Interface Modules (Option 03 or
Option 04) is installed in your CTS 710.
Appendix E: Incoming Inspection Test
CTS 710 SONET Test Set User Manual
E–5
2. Perform the initial setup of the CTS 710 with the following
sequence:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS RECALL
INSTRUMENT
SETUPS
DEFAULT
FACTORY
SETTINGS
Recall Setup
TRANSMIT TRANSMIT
SETTINGS
Tx/Rx Settings Coupled
POINTERS &
TIMING
Pointer Control Set Value
Pointer Value set
to
Default 522
3. To check the STS-1 rate, press the CLEAR HISTORY button,
wait two seconds, and then verify that no yellow history lights are
on.
4. To check the STS-3 rate, set the transmit rate as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TRANSMIT TRANSMIT
SETTINGS
Transmit Rate STS-3
NOTE. When changing the transmit rate, the LOP, LOS, and LOF
history lights may turn on; this is normal. However, the red event
lights should not stay on.
5. Press the CLEAR HISTORY button, wait two seconds, and then
verify that no yellow history lights are on.
6. Steps 7 through 10 of this procedure apply only if one of the
Electrical/Optical Plug-in Interface Modules (Option 03 or
Option 04) is installed in your CTS 710.