Tektronix cts710 Surge Protector User Manual


 
Setting TributaryPDH Receive Parameters
3–122
CTS 710 SONET Test Set User Manual
Setting the Receive Rate
To set the receive rate:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE RECEIVE
SETTINGS
(see Figure 3–46)
Receive Rate as appropriate
H Select DS1, DS3, or DS3–>DS1 Demux to receive a tributary
rate signal.
Selecting DS3–>DS1 Demux will demultiplex a DS1 signal from
a received DS3 signal. All measurements are made at the DS1
level except for Loss of Signal.
Available DSn
Receive Rates
Figure 3–46: Tributary Signal Receive Rates
Setting TributaryPDH Receive Parameters
3–122
CTS 710 SONET Test Set User Manual
Setting the Receive Rate
To set the receive rate:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE RECEIVE
SETTINGS
(see Figure 3–46)
Receive Rate as appropriate
H Select DS1, DS3, or DS3–>DS1 Demux to receive a tributary
rate signal.
Selecting DS3–>DS1 Demux will demultiplex a DS1 signal from
a received DS3 signal. All measurements are made at the DS1
level except for Loss of Signal.
Available DSn
Receive Rates
Figure 3–46: Tributary Signal Receive Rates