Theory of Operation
Analog Section Detailed Circuit Description
2
2-143
connector is connected back to the OUTPUT HI SENSE and OUTPUT LO SENSE
binding posts.
Attenuation is calibrated starting with 70 dB and decreased in 10 dB steps to 0 dB. The
Switch Matrix assembly connects the output of the Wideband AC module to its internal
cal zero amplifier circuit during all attenuator calibrations except the 0 dB and 10 dB.
The internal cal zero amplifier circuit provides a gain of 10 to Wideband output during
20 dB to 70 dB attenuation calibration. The output of this amplifier is connected to the
RCL line which is routed to the +input of the adc circuit on the DAC assembly. Output
from the DAC is connected to the -input of the adc circuit and adjusted until a null is
achieved. The exact attenuator value is then determined.