Fluke 5720A Power Supply User Manual


 
Calibration and Verification
Full Verification
3
3-37
9. Change the Calibrator frequency to 20 Hz, 40 Hz, 1 kHz, 5 kHz and 10 kHz. At each
frequency record the error display on the 5790A in Table 3-28 or 3-29. Verify that
the results are within limits shown.
10. Repeat steps 2 through 9, but replace the 200 metal film resistor with the 2 k
resistor, and use 200 µA instead of 2 mA.
3-25. Rationale for Using Metal-Film Resistors to Measure AC Current
To be able to measure alternating current, a system comprised of a suitable ac shunt and
ac detector is required. First let us consider the ac shunt. For this example we will use a 2
k metal film resistor. At frequencies up to 10 kHz, the equivalent circuit of the resistor
can be illustrated as in Figure 3-14. Values typical for shunt capacitance and series
inductance are 2 pF (Cs) and 0.01 µH (Ls). For comparison, wire has approximately 0.02
µH/inch. At 10 kHz, the reactance of Cshunt is 8 M, and the reactance of Lseries is 0.6
m. The formulae to use are:
SHUNT
C
SERIES
L
R
F3-10.EPS
Figure 3-14. Metal Film Resistor Equivalent Circuit
(1/Z)
2
= (1/R)
2
+ (1/XC)
2
(1)
(Z)
2
= (R)
2
+ (XL)
2
(2)
Where R = resistance Xc = Capacitive Reactance
Z = network impedance XL = Inductive Reactance
We can see that these effects can be ignored, because their contribution to errors in the
measurement process is less than 1 ppm. That is, the metal film resistor’s self reactance
is totally dwarfed by the reactance of the measuring circuit, which is overwhelmingly
capacitive.
If a detector as shown in Figure 3-15 has an input impedance of 10 M shunted by 123
pF, then the effects of Xc must be accounted for. We can ignore the net resistance
change introduced by the 10 M detector resistance.
SHUNT
R
DET.UUT
F3-11.EPS
Figure 3-15. Metal Film Resistor in Test Circuit