Fluke 5720A Power Supply User Manual


 
Troubleshooting
Component-level Troubleshooting
5
5-55
These relays are latched in the set or reset position and remain in that position even
after power has been removed. To troubleshoot the signal path through the Switch
Matrix assembly simply set the Calibrator to the function in which a problem is
detected and then turn off the Calibrator. Now remove the Switch Matrix from the
instrument and, using an ohmmeter, check the signal path by using the relay chart in
the schematic section. Check individual relays by connecting an external power
supply (a typical 9V battery works well) across the set coil to place the relay in the
set position or the reset coil to place the relay in the reset position.
1 V
10 ms
F5-15.EPS
Figure 5-15. Waveform at TP16
2. Check the Motherboard relay RLY+V supply voltage. Set the Calibrator to 10V dc,
operate. Connect an oscilloscope to RLY+V (common to TP9),which is at the
cathode of CR2, and verify it is 4.5V ±0.5V. Now verify that each time the EX SNS
key is pressed the voltage at RLY+V momentarily jumps to 7.5V ±1.2V. If a failure
is detected, check the control line PC5, U17, Q1, Q2, and associated components.
3. Check the Motherboard relay control lines. Motherboard relays K1-K10 are
controlled by U13 on the Switch Matrix assembly. If aMotherboard relay problem is
suspected, verify U13 by using the relay chart in the Switch Matrix schematic
section and checking for correct logic levels at its output.
4. Check the 2.2V Range Attenuator. Set the Calibrator to 2V dc, operate. Using a
DMM measure the dc voltage at TP2 (common to TP8) and verify it is 2V ±6 mV.
Note
This 2V can be as much as ±0.12V if the DAC assembly is uncalibrated. If
a failure is detected, check K1, K2, and the 5:1 divider part of the HR1
resistor network.
5. Check the 2.2V Range Amplifier. Set the Calibrator to 2V dc, operate. Using a
DMM measure the dc voltage at TP4 (common to TP8) and verify it is 2V ±6 mV.