Calibration and Verification
Determining Test Limits for Other Calibration Intervals
3
3-67
Table 3-32. Wideband Flatness Test Record (cont)
Output
Level
Frequency
Measured
Flatness Error
Spec
± ppm
(90-day)
5790A
1-Year
Spec
TUR GB
Factor
Test Limit
± ppm
30 mV 10 kHz 1000 500 2.00 0.932 932
120 kHz 1000 500 2.00 0.932 932
500 kHz 1100 500 2.20 0.945 1040
2 MHz 1100 500 2.20 0.945 1040
5 MHz 2100 1000 2.10 0.939 1972
10 MHz 2100 1000 2.10 0.939 1972
20 MHz 4100 1500 2.73 0.972 3985
30 MHz 10100 3500 2.89 0.977 9868
10 mV 10 kHz 1000 500 2.00 0.932 932
120 kHz 1000 500 2.00 0.932 932
500 kHz 1300 700 1.86 0.920 1196
2 MHz 1300 700 1.86 0.920 1196
5 MHz 2300 1000 2.30 0.950 2185
10 MHz 2300 1000 2.30 0.950 2185
20 MHz 4300 1700 2.53 0.962 4137
30 MHz 10300 3700 2.78 0.972 10012
3 mV 10 kHz 1000 500 2.00 0.932 932
120 kHz 1000 500 2.00 0.932 932
500 kHz 2000 1033 1.94 0.928 1856
2 MHz 2000 1033 1.94 0.928 1856
5 MHz 4000 1333 3.00 0.980 3920
10 MHZ 4000 1333 3.00 0.980 3920
20 MHz 6000 2033 2.95 0.978 5868
30 MHz 16000 4033 3.97 0.999 15984
1 mV 10 kHz 1000 500 2.00 0.932 932
120 kHz 1000 500 2.00 0.932 932
500 kHz 5000 1700 2.94 0.978 4890
2 MHz 5000 1700 2.94 0.978 4890
5 MHZ 7000 2700 2.59 0.965 6755
10 MHz 7000 2700 2.59 0.965 6755
20 MHz 9000 4000 2.25 0.948 8532
30 MHz 30000 9000 3.33 0.988 29640