Fluke 5720A Power Supply User Manual


 
5700A/5720A Series II Calibrator
Service Manual
1-16
Table 1-4. DC Voltage Secondary Performance Specifications and Operating Characteristics
Temperature Coefficient
Adder [Note 2]
Noise
Range Stability [Note 1]
± 1 °C
10 °-40 °C 0 °-10 °C
and
Linearity
± 1 °C
Bandwidth
0.1-10 Hz
Bandwidth
10 Hz-10 kHz
24 Hours 40°-50 °C pk-pk RMS
± (ppm output + µV) ± (ppm output + µV)/°C ± (ppm output + µV) µV
220 mV
2.2 V
11 V
22 V
220 V
1100 V
0.3 + 0.3
0.3 + 1
0.3 + 2.5
0.4 + 5
0.5 + 40
0.5 + 200
00.4 + 0.1
00.3 + 0.1
0.15 + 0.2
00.2 + 0.4
00.3 + 5
00.5 + 10
1.5 + 0.5
1.5 + 2
0 1 + 1.5
1.5 + 3
1.5 + 40
0 3 + 200
0 1 + 0.2
0 1 + 0.6
0.3 + 2
0.3 + 4
0 1 + 40
0 1 + 200
0.15 + 0.1
0.15 + 0.4
0.15 + 2
0.15 + 4
0.15 + 60
0.15 + 300
5
15
50
50
150
500
Notes:
1. Stability specifications are included in the Absolute Uncertainty values in the primary specification
tables.
2. Temperature coefficient is an adder to uncertainty specifications that does
not
apply unless operating
more than ±5 °C from calibration temperature.
Minimum output: 0 V for all ranges, except 100 V for 1100 V range
Maximum load: 50 mA for 2.2 V through 220 V ranges; 20 mA for 1100 V range; 50 output impedance
on 220 mV range; all ranges <1000 pF, >25
Load regulation: <(0.2 ppm of output + 0.1ppm of range), full load to no load
Line regulation: <0.1 ppm change, ± 10 % of selected nominal line
Settling time: 3 seconds to full accuracy; + 1 second for range or polarity change; + 1 second for 1100 V
range
Overshoot: <5 %
Common mode rejection: 140 dB, DC to 400 Hz
Remote sensing: Available 0 V to ±1100 V, on 2.2 V through 1100 V ranges