5700A/5720A Series II Calibrator
Service Manual
3-70
Table 3-37. AC V Distortion Test Summary
5700A/5720A Series
II Output
Load Resistors Frequency Measured
Distortion
Max.
Distortion
2V 100Ω, 1/8W 10 Hz, 20 Hz 0.054%
1 kHz, 20 kHz, 50 kHz, 0.044%
100 kHz, 200 kHz, 500 kHz 0.355%
20V 1 kΩ, 1/2W 10 Hz, 20 Hz 0.0535%
20 kHz, 100 kHz 0.0385%
200 kHz, 500 kHz 0.304%
200V 10 kΩ, 5W 10 Hz, 20 Hz 0.055%
50 kHz, 100 kHz 0.1065%
300V(1) 15 kΩ, 5W 40 Hz 0.1%
300V(1) 15 kΩ, 5W 50 kHz 0.3%
300V(1) 15 kΩ, 5W 70 kHz 0.4%
(1) The 5700A/5720A Series II maximum volt-Hertz product is (2.2 x 10
7
) The 300V level assumes that a
Fluke 5725A Amplifier is attached.
Table 3-38. Test Record for Flatness Check of the AC 2 mV Range
Frequency UUT Error Display Reading Limits
10 Hz
20 Hz ±0.26%
40 Hz
1 kHz ±0.26%
20 kHz ±0.29%
50 kHz ±0.49%
100 kHz ±0.87%
300 kHz
500 kHz ±1.83%
1 MHz