Calibration and Verification
Introduction
3
3-3
3-1. Introduction
This chapter gives procedures for 5700A/5720A Series II calibration, verification,
acceptance testing, and performance testing. Information here applies to testing the
performance of and calibrating a normally operating 5700A/5720A Series II. In case of
malfunction, refer to Chapter 5, Troubleshooting, which explains how to use self
diagnostic tests to identify a faulty module. Calibration and Performance Testing is
presented in the following three parts:
• Calibration, which is to be done at the beginning of every calibration cycle. This is
the same procedure as in Chapter 7 of the 5700A/5720A Series II Operator Manual.
It uses three external standards; 1Ω, 10 kΩ, and 10V dc. The procedure is repeated
here for convenience. Also included in this part are procedures for doing Calibration
Check and Range Calibration.
• Full Performance Verification, which is the full verification procedure,
recommended every two years. Part of this procedure is Wideband AC Module
(Option 5700A-03) flatness calibration, also recommended only every two years.
Optional Tests, which are recommended following repair or for use in acceptance
testing. These tests include such checks as load regulation, noise, and distortion. These
tests are not required on a routine basis. They are not necessary after a Calibrator passes
Full Performance Verification.