Fluke 5720A Power Supply User Manual


 
Troubleshooting
Interpreting Diagnostic Fault Codes
5
5-23
assembly. The resulting 0.2V across the 10 k portion of Z5 is connected to the +input
of the differential amplifier. The output of the differential amplifier is routed to the adc
circuit on the DAC assembly, via the RCL line, and measured for 0V ±3% ±15 mV. A
failure indicates the calibration of the 100 resistance on the Ohms Main assembly
(A10) is probably at fault.
3341 A9/A10: 190 Ohm Cal Diag Fault
The 2/5/10V SOURCE circuit is set for 2V and connected across the 1.9 k string on the
Ohms Main assembly. The resulting 0.2V across the 190 portion is connected to the -
input of the Differential Amplifier on the Ohms Cal assembly. The output of the DAC
assembly is set to 2V and connected across the 100 k resistance in Z5 on the Ohms Cal
assembly. The resulting 0.2V across the 10 k portion of Z5 is connected to the +input
of the differential amplifier. The output of the differential amplifier is routed to the adc
circuit on the DAC assembly, via the RCL line, and measured for 0V ±3% ±15 mV. A
failure indicates the calibration of the 190 resistance on the Ohms Main assembly
(A10) is probably at fault.
3400 A12: 8255 Control Word
The diagnostics software reads the control word of the 8255 IC on the Oscillator Control
assembly. A failure indicates the Oscillator Control assembly (A12) is probably at fault.
3401 A12: 8255 Port A
The diagnostics software writes and reads data to port A of the 8255 IC on the Oscillator
Control assembly. A failure indicates the Oscillator Control assembly (A12) is probably
at fault.
3402 A12: 8255 Port B
The diagnostics software reads port B of the 8255 IC on the Oscillator Control assembly
during the instruments dormant state. A failure indicates the Oscillator Control assembly
(A12) is probably at fault.
3403 A12: 8255 Port C
The diagnostics software reads port C of the 8255 IC on the Oscillator Control assembly
during the instruments dormant state. A failure indicates the Oscillator Control assembly
(A12) is probably at fault.
3404 A13: 8255 Control Word
The diagnostics software reads the control word of the 8255 IC on the Oscillator Output
assembly. A failure indicates the Oscillator Output assembly (A13) is probably at fault.
3405 A13: 8255 Port A
The diagnostics software writes and reads data to port A of the 8255 IC on the Oscillator
Output assembly. A failure indicates the Oscillator Output assembly (A13) is probably at
fault.