Fluke 5720A Power Supply User Manual


 
Introduction and Specifications
AC Voltage Specifications
1
1-21
Table 1-9. AC Voltage Secondary Performance Specifications and Operating Characteristics
Stability
± 1 °C [Note 1]
Temperature Coefficient Maximum
Distortion
Range Frequency 24 Hours 10°-40 °C 0°-10 °C and
40°-50 °C
Output Impedance Bandwidth
10 Hz-10 MHz
Hz ± µV ±µV/°C Ω± (% output + µV)
2.2 mV
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
3
5
5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
50
00.05 + 10
0.035 + 10
0.035 + 10
0.035 + 10
0.035 + 10
000.3 + 30
000.3 + 30
000 1 + 30
22 mV
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
5
10
15
0.2
0.2
0.2
0.4
0.5
0.6
1
1
0.3
0.3
0.3
0.5
0.5
0.6
1
1
50
00.05 + 11
0.035 + 11
0.035 + 11
0.035 + 11
0.035 + 11
000.3 + 30
000.3 + 30
000 1 + 30
± (ppm output + µV) ± (ppm output µV)/°C
220 mV
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
080 + 15
012 + 2
010 + 2
010 + 2
020 + 4
100 + 10
200 + 20
002 + 1
002 + 1
002 + 1
015 + 2
015 + 4
080 + 5
080 + 5
080 + 5
02 + 1
02 + 1
02 + 1
15 + 2
15 + 4
80 + 5
80 + 5
80 + 5
50
00.05 + 16
0.035 + 16
0.035 + 16
0.035 + 16
0.035 + 16
000.3 + 30
000.3 + 30
000 1 + 30
Load Regulation
±(ppm output+ µV)
2.2 V
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
080 + 15
012 + 4
015 + 5
015 + 5
030 + 10
070 + 20
150 + 50
050 + 10
015 + 5
002 + 1
010 + 2
010 + 4
080 + 15
080 + 40
080 + 100
50 + 10
15 + 5
05 + 2
15 + 4
20 + 4
80 + 15
80 + 40
80 + 100
0010 + 2
0010 + 2
0010 + 4
0030 + 10
0120 + 16
0300 ppm
0600 ppm
01200 ppm
00.05 + 80
0.035 + 80
0.035 + 80
0.035 + 80
0.035 + 80
000.3 + 110
000.3 + 110
000 1 + 110
22 V
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
080 + 15
012 + 8
015 + 10
015 + 10
030 + 15
070 + 100
150 + 100
050 + 100
015 + 30
002 + 10
010 + 20
010 + 40
080 + 150
080 + 300
080 + 500
50 + 100
15 + 40
04 + 15
20 + 20
20 + 40
80 + 150
80 + 300
80 + 500
0010 + 20
0010 + 20
0010 + 30
0030 + 50
0080 + 80
0100 + 700
0200 + 1100
0600 + 3000
00.05 + 700
0.035 + 700
0.035 + 700
0.035 + 700
0.035 + 700
000.3 + 800
000.3 + 800
000 2 + 800
220 V
0010 - 20
0020 - 40
0040 - 20 k
020 k - 50 k
050 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 200
080 + 150
012 + 80
015 + 100
015 + 100
030 + 400
100 + 10,000
200 + 20,000
0 50 + 1000
0 15 + 300
00 2 + 80
0 10 + 100
0 10 + 500
0 80 + 600
0 80 + 800
0 80 + 1000
50 + 1000
15 + 300
04 + 80
20 + 100
20 + 500
80 + 600
80 + 800
80 + 1000
0010 + 200
0010 + 200
0010 + 300
0030 + .600
0080 + 3,000
0250 + 25,000
0500 + 50,000
1000 + 110,000
00.05 + 10,000
00.05 + 10,000
00.05 + 10,000
00.05 + 10,000
000.1 + 13,000
001.5 + 50,000
001.5 + 50,000
003.5 + 100,000
±(ppm output + mV) ±(ppm output)/°C ±(% output)
1100 V
0015 - 50
0050 - 1 k
150 + 0.5
020 + 0.5
50
2
50
5
0010 + 2
0010 + 1
0.15
0.07
Note: 1. Stability specifications are included in Absolute Uncertainty values for the primary specifications.